Test characteristics

  • High diagnostic sensitivity
  • Wide measuring range up to 8.7 μg FEU/mL
  • High antigen excess security up to 50 μg FEU/mL
  • Excellent precision already in the low range of the clinical cut-off
  • On-board stability up to 6 weeks
  • Calibration stability up to 6 weeks
  • Applicable to common clinical chemistry analyzers

Performance data

Precision

Intra-assay
(n=20)
Mean
[µg FEU/mL]
SD
[µg FEU/mL]
CV
[%]
sample 10.720.011.76
sample 21.020.011.35
sample 33.870.051.21
Inter-assay
(n=20)
Mean
[µg FEU/mL]
SD
[µg FEU/mL]
CV
[%]
sample 10.660.034.59
sample 20.950.022.18
sample 33.590.041.10